ME MB 3 ENDO

As evidenced by its ergonomic handling, the ME MB3 utilizes proven technology. Special rotary switches are provided to set the power individually for each operating mode, with the settings indicated on a digital display.

 

24
7
SUPPORT

As evidenced by its ergonomic handling, the ME MB 3 utilizes proven technology. Special rotary switches are provided to set the power individually for each operating mode, with the settings indicated on a digital display. The appropriate types of current can be selected easily via dedicated buttons arranged on the membrane keyboard.

  • Power requirements: 220–240 V, 50–60 Hz; or 100–127 V, 50–60 Hz (adjustable by Technical Service via plug-in jumper inside  unit)
  • Power input:
  • without HF output: approx. 26 VA
  • at max. power output: approx. 800 VA
  • Protection class: I
  • Classification acc. to MDD: II b
  • Leakage currents: LF and HF in compliance with IEC 601, Part 2-2
  • Type: CF; defibrillation-proof
  • Nominal frequency: 450 kHz
  • Modulation frequency: 30 kHz

HF OUTPUT POWER:

Type of current

  • Cut 1 (pure)
    • Power: max. 400 W at 300 Ω
    • Crest factor: 6 at 300 Ω
    • Voltage: max. 2,300 Vpp
  • Cut 2 (blend)
    • Power: max. 300 W at 300 Ω
    • Crest factor: 1.9 at 300 Ω
    • Voltage: 2,500 Vpp
  • Cut 3 (super blend)
    • Power max. 200 W at 300 Ω
    • Crest factor: 2.4 at 300 Ω
    • Voltage: 2,700 Vpp
  • Forced Prep
    • Power: max. 120 W at 200 Ω
    • Crest factor: 4.9 at 200 Ω
    • Voltage: 3,900 Vpp
  • Endo Mode
    • Power: max. 100 W at 200 Ω
    • Crest factor: 1.6 at 200 Ω
    • Voltage: 2,800 Vpp
  • Contact coagulation
    • Power: max. 250 W at 200 Ω
    • Crest factor: 3.4 at 200 Ω
    • Voltage: 3,200 Vpp
  • Spray coagulation
    • Power: max. 120 W at 300 Ω
    • Crest factor: 5.6 at 300 Ω
    • Voltage: 9,000 Vpp
  • Bipolar coagulation
    • Power: max. 100 W at 100 Ω
    • Crest factor: 2.1 at 100 Ω
    • Voltage: 600 Vpp
  • Duty type: intermittent 10 s / 30 s = duty factor of 25%
  • Line fuses:
  • 220–240 V: T 4 A (slow-blow)
  • 100–127 V: T 8 A (slow-blow)
  • Sound level:
  • HF current activation:
  • 40 - 65 dB (adjustable)
  • Alarm: 65 dB(A)
  • Weight: 6 kg
  • Interference suppression:
  • Limits in conformity with EN 55011
  • Interference immunity in conformity with IEC 801
  • Dimensions
  • Width: 405 mm; height: 135 mm; depth: 380 mm
  • CE-marking according to 93/42/EEC

ME_MB3_En