espa2
espa
English
BACK TO PRODUCTS
POWERCUBE® BODY+
SPECIALTIES : Pulmonology
The body plethysmography is a very sensitive lung measurement used to detect complex or mixed lung pathology that might be missed with conventional pulmonary function tests.
DETAILED DESCRIPTION
TECHNICAL SPECIFICATIONS
ACCESSORIES
RELATED FILES
VIDEO

Upmost long-term stability in conjunction with our new analysis algorithms enables you to initially differentiate between asthma, COPD, restriction or any combination of these disorders just by looking at the shape of the curve.

  • Ultrasonic flow measurement as standard
  • Calibration free
  • Easily adjustable patient interface
    • The choice of either a broad bench seat (max. 160 kg) or a height-adjustable swivel chair (max. 120 kg)
    • Automatic and simultaneous calibration of both the cabin and the mouth pressure at three different frequencies
  • BodyLiveCal to test your TGV
  • Ceiling lighting that can illuminate in different colours, including white
  • LEDs indicating the locked state of the door

Flow-Measurement

Principle Time

Range 0 to ± 20 L/s

Accuracy ± 2%, 50 ml/s (whichever is greater)

Resistance <0.05 kPa/(l/s) at 14 l/s

Volume-Measurement

Principle Digital Integration

Range Not limited

Accuracy ± 2% or 0.05 L

Mouth Pressure

Principle Piezo resistive

Range ± 30 kPa

Accuracy ± 1%

CO Analyzer

Principle Non-dispersive Infrared Analyzer

Range 0 – 0.5%

Accuracy ± 1%

He Analyzer

Principle Ultrasound

Range 0 – 20 % He

Accuracy ± 1%

Cabin Pressure

Principle Piezo resisitive

Range ± 1 kPa

Accuracy ± 1%

Weight

Weight of cabin 150 kg

Dimensions

W/o door Handle 86 cm x 185 cm x 71 cm